Home Up Audiometers Audiometric Test Rooms Electroacoustic Ears Electrocardiographs Noise Dosimeters Services Software Sound Level Meters Vision Testers
| |
|
"Flowmate
V+" by Spirometrics
|
Features |
|
The
FlowmateV plus features real time graphics,
simultaneously displaying the flow/volume loop and the volume/time
curve allowing for immediate review.
Contamination
Control. Cross-contamination control via an external
filter system.
Worldwide
Warranty. Full 12-month warranty of the Flowmate V+'s
components, parts and labor. Additional 3-year Extended Warranty /
Trade-in program available.
|
Dimensions |
|
Height |
5.5" |
|
Depth |
9.0" |
|
Width |
11.375" |
Weight |
|
Net
Weight |
10.5
lbs |
Power
Requirements |
|
Voltage |
100-240
VAC, 50-60Hz |
|
Amperage |
1.0A
to 500 Milliamps (As supplied to the external power supply) |
|
Wattage |
45
Watts maximum |
Power
Connection |
|
IEC standard
three prong connector with ground. Permits wide variety of
international primary power connector cord sets. Supplied with
hospital grade plug USA and Canada. |
|
Environmental |
|
Temperature
Operating Range |
50F
to 104F (10C to 40C) |
|
Maximum
Storage Temperature |
150F
(68C) |
User
Interface |
|
Display |
256
by 128 graphic LCD Display |
|
Keypad |
26
key membrane switch keypad with tactile feedback. |
Patient
Data |
|
ID
Number |
Up to
16 digits including periods (.), and slashes (/). |
|
Age |
1 to
99 years. |
|
Sex |
Male,
Female. |
|
Race |
Caucasian,
Black, Non-Caucasian, Hispanic, Asian. |
|
Height |
Up to
three digits, in inches or centimeters. |
|
Weight |
Up to
three digits (optional entry) customizable as lbs. or kgs. |
|
Smoke |
0 -
60 cigarettes per day. |
TEST
MANEUVERS |
|
FVC |
Forced
Vital Capacity |
|
MVV |
Maximum
Voluntary Ventilation |
|
VC |
Vital
Capacity (otherwise known as Slow VC) |
|
Challenge |
Methacholine
Challenge |
INDICES
CALCULATED |
|
FVC
Maneuver |
|
|
FVC |
Forced
Vital Capacity |
|
|
FEV1 |
1
second Forced Vital Capacity |
|
|
FEV3 |
* 3
second Forced Vital Capacity |
|
|
FEV1/FVC |
% FEV1
Ratio |
|
|
FEF25-75% |
Mid
Expiratory Flow Rate |
|
|
FEF75-85% |
*
Late Expiratory Flow Rate |
|
|
FEF200-1200 |
*
Forced Expiratory Flow Rate between 200ml and 1200ml |
|
|
FEF50 |
*
Forced Expiratory Flow Rate at 50% of FVC |
|
|
PEF |
Peak
Expiratory Flow |
|
|
FIVC |
*
Forced Inspiratory Vital Capacity |
|
|
FIF50 |
*
Forced Inspiratory Flow at 50% of FVC |
|
|
FIF50/FEF50 |
*
Ratio of FEF50 to FIF50 |
|
|
PIF |
*
Peak Inspiratory Flow |
|
|
COPD |
Risk
Assessment |
|
|
Lung
Age |
Lung
Age Assessment based on FEV1 |
|
NOTE:
FEV timing is based on back extrapolation
* these indices appear on the expanded reports |
|
|
|
|
|
|
|
|
|
MVV
Maneuver |
|
|
MVV
Volume |
Volume
Expired per minute |
|
|
MVV
Rate |
Breaths
per minute |
|
|
|
|
|
|
|
|
|
VC
Maneuver |
|
|
VC |
Vital
Capacity |
|
|
ATI |
Air
Trapping Index |
|
|
|
|
|
|
|
|
|
Challenge |
|
|
FVC |
Forced
Vital Capacity |
|
|
FEV1 |
1
second Forced Vital Capacity |
|
|
PD20 |
Provocative
Dose |
FLOW
SENSOR |
|
Maximum
Flow Rate |
+/-
16liters/second |
|
Accuracy |
+/-
3% |
|
Resolution |
12
bit signed A/D |
|
Sampling
Rate |
150Hz |
|
Back
Pressure |
1.1
cmH2O/L/s@ 14L/S |
BTPS
calculated automatically - based on user input for environmental
conditions. |
|
Temperature |
50F to 104F (10C to 40C) |
|
Barometric Pressure |
20 to 35 inches of mercury, 500 to 900
millimeters of mercury. Input values determines units
automatically. |
BEST
TEST DETERMINATIONS (pertains to FVC trials only) |
|
FVC |
Largest
regardless of test |
|
FEV1 |
Largest
regardless of test |
|
Ratio |
Combination
of Best FEV1/Best FVC |
|
Flow
Rates |
Taken
from the trace with the greatest sum of FVC and FEV1 |
|
Patient
Effort |
Requires
clinician to evaluate acceptable or unacceptable patient effort.
Discards tests the clinician deems as unacceptable. |
|
Variability |
Compares
indices of current test with best test in memory. Displays and
prints variability as % or as absolute difference (user selectable
in customization). Variability greater than 5% or 0.2L (depending
on user selection) for the FVC and/or the FEV1 is regarded as poor
reproducibility. |
|
Expiratory
Time |
If
forced expiratory time is less than 6 seconds, a "TEST TOO
SHORT" message will be displayed. |
|
Extrapolated
Volume |
If
extrapolated volume is greater than 5% of the FVC, a "SLOW
START" message will be displayed. |
|
Interpretations |
Obstruction,
Restriction, or Ventilary disorder. |
INPUT/OUTPUT
COMMUNICATION PORTS |
|
Graphic
Output |
Printouts
according to ATS scale specifications when user selects option in
customization. |
|
Printer
Options |
Internal
thermal printer using 112mm wide thermal paper. External printer
with selectable print drivers within customization. Standard DB-25
Female IBM PC pin out to standard Centronic connection. |
|
Patient
Storage |
Internal
minimum of 7 patients (variable depending on number of FVC tests
performed and saved). External storage on 1.44 Mb formatted disk
minimum 45 patients (variable depending on number of FVC tests
performed and saved).
|
FVC |
-
6 trials each patient |
|
MVV |
-
2 trials each patient |
|
VC |
-
6 trials each patient |
|
|
Test
Numbering |
Numbers
each patient trial that is acceptable consecutively regardless of
the type of test for recall later. |
INSTRUMENT
CALIBRATION |
|
Syringe |
Selectable
from 1 to 9 liters at ATPS. Daily calibration is recommended on
days testing is to be performed. Calibration reports print out
automatically if printer is connected and is on-line. |
|
|