[Company Logo Image]  

Home Up News Feedback Contents Search Technical Tips


Spirometrics Flowmate V Plus


Home
Up
Audiometers
Audiometric Test Rooms
Electroacoustic Ears
Electrocardiographs
Noise Dosimeters
Services
Software
Sound Level Meters
Vision Testers

 

 


"Flowmate V+" by Spirometrics
 
Flowmate V+
A Fully Automated Spirometer
Features
The FlowmateV plus features real time graphics, simultaneously displaying the flow/volume loop and the volume/time curve allowing for immediate review.

Contamination Control. Cross-contamination control via an external filter system.

Worldwide Warranty. Full 12-month warranty of the Flowmate V+'s components, parts and labor. Additional 3-year Extended Warranty / Trade-in program available.


Dimensions
Height 5.5"
Depth 9.0"
Width 11.375"

 

Weight
Net Weight 10.5 lbs

Power Requirements
Voltage 100-240 VAC, 50-60Hz
Amperage 1.0A to 500 Milliamps (As supplied to the external power supply)
Wattage 45 Watts maximum

Power Connection
IEC standard three prong connector with ground. Permits wide variety of international primary power connector cord sets. Supplied with hospital grade plug USA and Canada.

Environmental
Temperature Operating Range 50F to 104F (10C to 40C)
Maximum Storage Temperature 150F (68C)

User Interface
Display 256 by 128 graphic LCD Display
Keypad 26 key membrane switch keypad with tactile feedback.

Patient Data
ID Number Up to 16 digits including periods (.), and slashes (/).
Age 1 to 99 years.
Sex Male, Female.
Race Caucasian, Black, Non-Caucasian, Hispanic, Asian.
Height Up to three digits, in inches or centimeters.
Weight Up to three digits (optional entry) customizable as lbs. or kgs.
Smoke 0 - 60 cigarettes per day.

TEST MANEUVERS
FVC Forced Vital Capacity
MVV Maximum Voluntary Ventilation
VC Vital Capacity (otherwise known as Slow VC)
Challenge Methacholine Challenge

INDICES CALCULATED
FVC Maneuver
FVC Forced Vital Capacity
FEV1 1 second Forced Vital Capacity
FEV3 * 3 second Forced Vital Capacity
FEV1/FVC % FEV1 Ratio
FEF25-75% Mid Expiratory Flow Rate
FEF75-85% * Late Expiratory Flow Rate
FEF200-1200 * Forced Expiratory Flow Rate between 200ml and 1200ml
FEF50 * Forced Expiratory Flow Rate at 50% of FVC
PEF Peak Expiratory Flow
FIVC * Forced Inspiratory Vital Capacity
FIF50 * Forced Inspiratory Flow at 50% of FVC
FIF50/FEF50 * Ratio of FEF50 to FIF50
PIF * Peak Inspiratory Flow
COPD Risk Assessment
Lung Age Lung Age Assessment based on FEV1
NOTE: FEV timing is based on back extrapolation
* these indices appear on the expanded reports
MVV Maneuver
MVV Volume Volume Expired per minute
MVV Rate Breaths per minute
VC Maneuver
VC Vital Capacity
ATI Air Trapping Index
Challenge
FVC Forced Vital Capacity
FEV1 1 second Forced Vital Capacity
PD20 Provocative Dose

FLOW SENSOR
Maximum Flow Rate +/- 16liters/second
Accuracy +/- 3%
Resolution 12 bit signed A/D
Sampling Rate 150Hz
Back Pressure 1.1 cmH2O/L/s@ 14L/S

BTPS calculated automatically - based on user input for environmental conditions.
Temperature 50F to 104F (10C to 40C)
Barometric Pressure 20 to 35 inches of mercury, 500 to 900 millimeters of mercury. Input values determines units automatically.

BEST TEST DETERMINATIONS (pertains to FVC trials only)
FVC Largest regardless of test
FEV1 Largest regardless of test
Ratio Combination of Best FEV1/Best FVC
Flow Rates Taken from the trace with the greatest sum of FVC and FEV1
Patient Effort Requires clinician to evaluate acceptable or unacceptable patient effort. Discards tests the clinician deems as unacceptable.
Variability Compares indices of current test with best test in memory. Displays and prints variability as % or as absolute difference (user selectable in customization). Variability greater than 5% or 0.2L (depending on user selection) for the FVC and/or the FEV1 is regarded as poor reproducibility.
Expiratory Time If forced expiratory time is less than 6 seconds, a "TEST TOO SHORT" message will be displayed.
Extrapolated Volume If extrapolated volume is greater than 5% of the FVC, a "SLOW START" message will be displayed.
Interpretations Obstruction, Restriction, or Ventilary disorder.

INPUT/OUTPUT COMMUNICATION PORTS
Graphic Output Printouts according to ATS scale specifications when user selects option in customization.
Printer Options Internal thermal printer using 112mm wide thermal paper. External printer with selectable print drivers within customization. Standard DB-25 Female IBM PC pin out to standard Centronic connection.
Patient Storage Internal minimum of 7 patients (variable depending on number of FVC tests performed and saved). External storage on 1.44 Mb formatted disk minimum 45 patients (variable depending on number of FVC tests performed and saved).
FVC - 6 trials each patient
MVV - 2 trials each patient
VC - 6 trials each patient
Test Numbering Numbers each patient trial that is acceptable consecutively regardless of the type of test for recall later.

INSTRUMENT CALIBRATION
Syringe Selectable from 1 to 9 liters at ATPS. Daily calibration is recommended on days testing is to be performed. Calibration reports print out automatically if printer is connected and is on-line.

 

 

Home ] Up ]

 

[ Home ] Audiometers ] Audiometric Test Rooms ] Electrocardiogrphs ] Noise Dosimeters ] Services ] Software ] Sound Level Meters ] Spirometers ] Vision Testers ]

 

Send mail to rdietz@lipindietz.com with questions or comments about this web site.
Copyright © 2001 Lipin/Dietz Associates, Inc.